MS-250 Mixed Signal Oscilloscope Option
Analog Pattern
1. To add the analog pattern to the digital pattern, leave your digital pattern as is and
skip to step 2.
To create an analog-only pattern, touch Set All To... and select Don't Care. This will
eliminate any meaningful digital pattern and activate all the Boolean operators.
2. Touch the Left Arrow button until the C1-EXT group of inputs is displayed in the main
section of the dialog.
3. Touch the Cx button for each input to be included in the pattern, and select whether it
must be High or Low compared to the threshold Level you will set.
Depending on your selection, a logical 1 (High) or 0 (Low) now appears on the dialog.
Leave Don't Care ("X") selected for any input you wish to exclude.
4. Select the Boolean operator (AND, NAND, OR, or NOR) that describes the relationship
among inputs (e.g., C1 must be High AND C2 must be Low).
NOTE:Only the AND operator is available when combining analog and digital patterns.
In the example above, all digital lines have been set to Don't Care ("X"), so all operators
are available.
5. Open the Levels dialog and enter the voltage threshold for each input included in the
trigger.
6. If you've included EXT as an input, open the Ext dialog and enter the Attenuation.